KF titrator (coulometric) · Calibration procedure
Water standard recovery
Injects a certified liquid water standard and checks that the coulometer finds the certified amount of water.
PURPOSE
Why this is checked
Frequency
Typically daily or before each series of samples, after changing reagent and after maintenance, as defined in your SOP.
Learn how the KF titrator (coulometric) worksYOU WILL NEED
Standards & tools
- Certified liquid water standard for coulometry, for example 1.0 mg/g (0.1%) or 0.1 mg/g, with certificate
- Calibrated analytical balance (readability 0.1 mg or better)
- Dry glass syringe with a long needle
BEFORE YOU START
Prerequisites
- Fresh anolyte (and catholyte, if used) in the cell, with enough remaining capacity
- Drift low and stable, for example below 10 µg/min
- Standard ampoule at room temperature and opened just before use
PROCEDURE
Step by step
- 1Open the ampoule, rinse the syringe with a small amount of standard and discard it.
- 2Draw up about 1 mL of standard, wipe the needle and weigh the filled syringe.
- 3Inject the standard through the septum below the anolyte surface and start the determination.
- 4Reweigh the empty syringe and enter the injected mass (difference).
- 5Repeat to obtain three determinations from the same ampoule.
- 6Calculate the recovery of each and the mean.
RECORD
Example observation table
| Injection | Mass injected (g) | Expected water (µg) | Water found (µg) | Recovery (%) |
|---|---|---|---|---|
| 1 | 1.0123 | 1012.3 | 1004.6 | 99.24 |
| 2 | 0.9876 | 987.6 | 982.9 | 99.52 |
| 3 | 1.0045 | 1004.5 | 1001.2 | 99.67 |
CALCULATION
How to calculate the result
Recovery (%) = Water found (µg) ÷ [Mass injected (g) × Certified content (µg/g)] × 100
- Certified content = 1.00 mg/g = 1000 µg/g
- Injection 1: expected = 1.0123 × 1000 = 1012.3 µg
- Recovery = 1004.6 ÷ 1012.3 × 100 = 99.24%
- Mean of three = 99.48%; %RSD = 0.22%
ACCEPTANCE CRITERIA
Pass if
- Mean recovery typically within 97.0–103.0% for a 1.0 mg/g standard (wider limits are often used for 0.1 mg/g standards), or the limits in your SOP.
- %RSD of replicate recoveries within the SOP limit, for example not more than 2%.
IF IT FAILS
What to do
- Check drift, septum and joints, then repeat with a fresh ampoule.
- Check the syringe weighing technique and the balance.
- Replace anolyte and catholyte and clean the generator electrode if recovery remains low.
- Do not test samples until a valid recovery is obtained; assess samples tested since the last passing check.